Polyimide coated fused silica
capillary tubing is a vital component in many separation science
techniques. In this application note we discuss methods for cutting
capillary tubing and show the results of different cutting
technologies.
Introduction
Fused silica capillary tubing’s
unique properties make it the material of choice for a number of
analytical techniques, including GC, CE, Capillary LC, & CEC.
Capillary is of growing importance in the area of Mass Flow Control
and Fluid Metering. Its dimensional precision is unmatched for
microfluidic interfacing.
Often the bulk capillary is delivered
in long lengths and is subsequently cut by the end user. For
products such as arrays, assemblies, and precisions pieces, the
capillary is often cut to specification by Polymicro. Regardless of
who performs this operation, the methodology for cutting and the
resulting end finish must be suitably matched to the application to
achieve optimum product performance.1 This note examines
the four most prominent methods for cutting capillary.
Experimental
The capillary
used in this study was TSP025375 (Polymicro, Phoenix, AZ). Standard
Cleaves were performed using a ceramic cleaving stone as described
previously.2 Precision Cleaved pieces were produced on a
proprietary cleaving device that is optimized for 363µm OD
capillary.
Saw Cut pieces
were made by bouling cleaved segments into a 20mm tube with a
proprietary polymer and then sawing on a Struers Accutom 50. Pieces
were then de-bouled and cleaned with Acetone.
Laser Cut pieces
were produced on a custom built, programmable laser station equipped
with a CO2 laser. A program specific to this capillary
size was written to control part processing.
Results
Figure 1a shows a
typical Standard Cleave, which has proven to be acceptable for a
range of applications, including CE and CEC. Good cleaves leave few
chips or cracks, and the polyimide should be flush with the
capillary end face. There is often a slight angle on the end face
and the surface can be uneven, as evident in the image.
Figure 1b
displays a Precision Cleaved capillary. Note the flatness across
the end-face when compared to Figure 1a. These types of low angle,
high quality cleaves are of particular interest in Capillary LC
connection lines and microfluidic interfacing, where unwanted chips,
cracks, and angles can increase system dead volume and
chromatographic band-broadening.
Figure 1c shows a
typical Saw Cut end face. Saw cutting is an efficient, bulk
production methodology. However, chipping can occur, end face
surfaces have a matte finish, and glass debris can contaminate the
i.d. Filling the i.d. with a wax or polymer reduces chipping and
introduction of debris, but must be removed after cutting; this can
be a limiting issue depending on the i.d. and part length. Saw cut
parts are often lapped and polished, significantly improving the
part end-face. Saw cutting and then polishing is the best method
for produce very short parts or parts with demanding length
tolerances.
Figure 1d
exhibits a Laser Cut capillary. Laser cutting is the method of
choice when a very smooth, defect free end-face is required. This
technique results in the removal of polyimide from the capillary o.d.
surface. It should be noted that the i.d. will remain free of
process debris.
Conclusion
This note
summarizes four commonly used techniques for cutting capillary.
Application requirements must always be considered. For assistance
with your specific application please contact a Polymicro Technical
Sales Specialist.
References
(1) J. Macomber and L. Begay, LCGC
Application Notebook, Sept. 2003 p.72.
(2) “Cleaving Procedure”, The Book
on the Technologies of Polymicro, Polymicro Technologies LLC
Publication, p A-2, (2005).